Description
JRCS JSTD-M15A
Technical Specifications
- Instrument type: Marine temperature indication controller
- Input: Pt100 RTD or thermocouple
- Measuring range: Selected by sensor
- Display: Pointer/digital temperature readout
- Output: Relay or 4 to 20 mA control output
- Setting: Panel pushbuttons/potentiometer
- Supply: AC24 V / DC24 V marine
- Accuracy: 1.0 class
- Temperature: -10 °C to +55 °C
- Storage: -25 °C to +70 °C
- Humidity: 5 % to 95 %, non-condensing
- Protection: IP55 (front panel)
- Installation: Panel flush-mount
- Certification: CCS
Functional Features
- Marine temperature monitoring and control in one unit
- Pt100 or thermocouple input
- Temperature limit alarm relay output
- Control output to temperature valves
- IP55 front panel resisting humid salty engine rooms
- Marine anti-vibration design
- Adjustable upper/lower alarm limits
- Real-time temperature display
- Marine wide-voltage supply
- Long-term reliable in harsh marine environment
Application Scenarios
- Ship engine room temperature monitoring
- Ship cargo hold and tank temperature monitoring
- Ship cooling water temperature control
- Ship fuel heater temperature control
- Offshore platform temperature monitoring systems
Materials and Construction
- Front panel: Corrosion-resistant metal panel with glass dial
- Housing: Die-cast metal with anti-corrosion paint
- Seal: Silicone rubber gasket
- PCB: FR4 board with conformal coating
- Mechanism: Pointer movement or digital display module
- Relay: Marine-grade electromagnetic relay
- Terminals: Marine waterproof terminals
Structural Features
- Panel flush-mount instrument structure
- Front panel integrates temperature dial, adjustment knob, and LEDs
- Rear housing terminals for wiring
- Gasket sealing panel-to-cabinet
- Internals vibration-hardened
- Marine metal housing anti-corrosion coating
Working Principle The Pt100 or thermocouple sensor signal enters input terminals; measurement circuitry converts sensor signal to temperature and drives the display mechanism. Comparison circuitry compares actual temperature to setpoint; exceeding limits triggers alarm relays. In control mode, output circuitry emits 4 to 20 mA proportional to deviation to control external valve or heater, maintaining temperature at setpoint.
Installation Requirements
- Panel cutout flush-mount on console
- Install gasket completely for IP55
- Connect sensor compensating wires per drawing
- Route control output to control valve
- Marine power supply
- Temperature -10 °C to +55 °C
- Calibrate temperature range after power-up
- Mount away from vibration and heat radiation
Precautions
- Sensor compensating wire polarity must be correct
- Do not short control output
- Periodically calibrate temperature accuracy
- Replace cracked front glass promptly
- Store spares in dry moisture-proof packaging
- Disconnect sensor and supply before service